ROSA T2/HT2 quantitative
Limit of detection < 10 µg/kg; no FGIS approval and no European claim.
| Analyte | Sum of T-2 and HT-2 |
| Toxin | T-2 and HT-2 toxin |
| Range | 0 – 250 / 0 – 2 500 µg/kg |
| Time | 10 min |
| Extraction | 70% methanol |
| Approval | Charm-validated |
| Conformance | 0 – 250 / 2 500 µg/kg |
| Method | solvent |
| Document | MRK-070 · brochure 2025 |
From sample to result
Six steps. The first three determine the outcome more than the strip itself, and they are the most difficult to correct afterwards.
- Sampling: 50 g of ground material, at least 90 % passing through a 20-mesh sieve. Record lot, silo and date before you start.
- Extraction: WET: one pouch of powder plus water according to the manual, shake for 1.5 minutes and allow to stand for 10 minutes. Methanol: 70 % methanol according to the manual.
- Dilute: First dilution for the normal range; the second only for levels above the midpoint of the range.
- Application: 100 µl extract onto the strip, strip fully inserted into the slot of the EZ-M. One strip per test channel.
- Incubate and read: 45 ± 1 °C, duration according to the test. The instrument reads automatically and checks volume, reuse, lens and door.
- Result: Quantitative value in µg/kg or mg/kg, with test ID, sample ID and operator. EZ Receiver compares it against the maximum level for the product.
Controls and storage
- Test and record negative and positive controls weekly; EZ Receiver tracks them with custom limits and a trend line.
- Store strips at 18–30 °C; discard remaining strips after 12 hours.
- Compare the incubator temperature display daily against 45 °C.
<a href="/en/alle-tests">Back to all tests</a>