ROSA WET-S5 Zearalenone quantitative

Limit of detection ≤ 25 µg/kg; FGIS compliance from 100 µg/kg.

AnalyteZearalenone
ToxinZearalenone
Range0 – 1 400 µg/kg
Time5 min
ExtractionWET powder + water
ApprovalFGIS: barley, maize, DDGS, sorghum, wheat
Conformance0 – 1 400 µg/kg
MethodWET
DocumentLF-ZEARQ-WETS5 · TKI 18-05-2023

From sample to result

Six steps. The first three determine the outcome more than the strip itself, and they are the most difficult to correct afterwards.

  1. Sampling: 50 g of ground material, at least 90 % passing through a 20-mesh sieve. Record lot, silo and date before you start.
  2. Extraction: WET: one pouch of powder plus water according to the manual, shake for 1.5 minutes and allow to stand for 10 minutes. Methanol: 70 % methanol according to the manual.
  3. Dilute: First dilution for the normal range; the second only for levels above the midpoint of the range.
  4. Application: 100 µl extract onto the strip, strip fully inserted into the slot of the EZ-M. One strip per test channel.
  5. Incubate and read: 45 ± 1 °C, duration according to the test. The instrument reads automatically and checks volume, reuse, lens and door.
  6. Result: Quantitative value in µg/kg or mg/kg, with test ID, sample ID and operator. EZ Receiver compares it against the maximum level for the product.

Controls and storage